THIS BLOG IS CREATED TO SERVE PEOPLE IN BEST POSSIBLE WAY
OFFERS SUGGESTIONS ON TRANSDUCER DESIGN ESPECIALLY ELECTROMECHANICAL TRANSDUCER ,INSTRUMENTATION,MEDICAL ELECTRONICS,ECG RELATED INFORMATION,NEW PROJECTS IN TRANSDUCER CALIBRATION SYSTEMS COULD BE UNDERTAKEN.GUIDANCE IN ISO 9001:2008 CERTIFICATION AND NBA ACCREDITATION PROCESS.
GAJANAN
SUSWAGATAM
OM NAMO GAJANANNAY NAMO GAURI ANANDANAY NAMO GAURIHRHIDAY NANDANAY NAMO SADSHIVA NANDANAY NAMASTE LAMBODARAY NAMO VGHNAHARAY NAMO MORESHWARAY NAMASTAYE CHINTAMANI NAMO MAHAGANAPATI NAMASTE GIRIJA NANDANAY NAMASTE BALLALAYA NAMO NAMAH VARADMURTAYE NAMAO SIDDHI VINAYAKAY NAMO DATTATREYAY NAMO AGASTI NAMO RENUKA MATAY
I WELCOME ALL THOSE WHO NEED HELP IN STUDIES IN INSTRUMENTATION,BIOMEDICAL INSTRUMENTATION,MECHATRONICS,TRANSDUCER DESIGNS,OPTCAL FIBER COMMUNICATION AND SENSOR DESIGNS,ANALOGUE COMMUNICATION,ELECTRONIC MEASUREMENT SYSTEMS,STUDIES IN PHYSICS AT 10TH AND TWELFTH LEVEL
BE A SCIENCE LOVER
BE AN ENGINEER ALWAYS
LOVE THY PROFESSION ALWAYS
LOVE YOUR CHOSEN BRANCH OF STUDIES AND DEVOTE YOUR LIFE TO DEVELOPE TECHNOLOGY FOR INDIA AND MANKIND
BE ALWAYS READY TO DO INTUITIVE ,INNOVATIVE REAEARCH
ALONG WITH MY PhD FRIENDS I WILL ALWAYS CONTRIBUTE TO DEVELOPE NEW TECHNOLOGY TO HELP CHILDREN,ELDERS ,WOMEN AND POOR OF THE WORLD TO FULL FILL TARGET OF UNITED NATIONS AND INDIA
ALWAYS REMEMBER THAT WASUDHEIWA KUTUMBAKAM
GANAPATI
GANAPATI BAPPA M0RAYA
MAHUR RENUKA MATA
KULASWAMINI RENUKA MATA
MAHUR VISIT
JAAI AND ARUNA AT KAAL BHAIRAWA MANDIR
MAHALAKSMI
KOLHAPOOR MAHALAKSHMI
Sapta Shrungi
SAPTA SHRUNGI MATA
TULAJA BHAWANI MATA
MAHARASHTRA WARDAYEENI
RENUKA MATA
RENUKA MATA MAHUR
Tuesday, August 18, 2009
ECG MACHINE TESTS
THE ECG MACHINE CAN BE TESTED USING SIMULATED ECG THIS HELPS STUDENTS OF ENGINEERING,WHO PURSUE A COURSE IN BIOMEDICAL ELECTRONICS,AT BE LEVEL IN THE UNIVERSITY AFFILIATED COLLEGES, TO TEST THE MACHINES MORE EFFECTIVELY AND PERFORM EXPERIMENTS BETTER.
THE TEST NEEDS FOLLOWING SIMULATED WAVEFORMS
.STANDARD PATTERN ECG AS PER NORMS WITH "R to R" REPETITION RATE OF 72 bpm OR "R to R" DISTANCE OF 800mS
ABSENCE OF "P" WAVE
ABSENCE OF "R" WAVE
ELEVATED"QRS COMPLEX"
"R " WAVE ONLY INVERTED
ABSENCE OF "T" WAVE
ELEVATED "ST" SEGMENT
CREATE "U" WAVE
FOLLOWING TESTS ON A "PATIENT" STUDENT" COULD BE CARRIED OUT SAFELY TO UNDERSTAND FUNCTIONING OF THE EC G MACHINE AND LEARN TO DIAGNOSE THE FAULTS. NO HARM TO THE STUDENT.
PUT RA LA RL LL LEADS AT APPROPRIATE LOCATIONS
SWITCH TO LEAD I POSITION ON THE INSTRUMENT/ECG MACHINE
RECORD WAVE FORM NOTE AT LEAST 4 "R" WAVES.
SWITCH TO LEAD II AND THEN TO LEAD III AND RECORD WAVE FORM
THEN INTERCHANGE RA AND RL POSITIONS AND RECORD WAVE FORMS
THEN INTERCHANGE RA & LL LEADS POSITIONS AND RECORD WAVE FORMS
INTERCHANGE BETWEEN LA AND LL LEADS AND RECORD WAVEFORMS
REMOVE RL AND RECORD FOR LEAD I,II,III ONLY.
INTERCHANGE RL AND LL AND RECORD
THE OBSERVATION OF ALL RECORDS OF THE WAVE FORMS TEACHES THE STUDENT ABOUT INCORRECT POSITIONS AND TEACHES ABOUT FAULTS.
PROFESSOR(ELECTRONICS) DEPARTMENT OF E & TC